User's Manual

4-30
Rad-87 Pulse CO-Oximeter Operator’s Manual
4
Operation
Setup Menu Level 3, continued
RRa Sensor Status Notifications
BUTTONS SETTINGS
Press the
Enter Button
again to
move to the
next menu.
On
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and move
to the next menu option.
OR
press the Display Button to exit without saving the new
setting and to return to the home display screen.
Off
(Default)
Rapid Desat Limit
The Rapid Desat Limit is designed to detect rapid desaturations of 5% or 10% below the low
alarm limit and overrides the Alarm Delay feature when activated.
BUTTONS SETTING
Press the Enter
Button again
to move to the
next menu.
5%
(Default)
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and move
to the next menu.
OR
press the Display Button to exit without saving the new
settings and to return to the home display screen.
Off
10%
Alarm On/Off
BUTTONS SETTING
Press the Enter
Button again
to move to the
next menu.
On
(Default)
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and move
to the next menu.
OR
press the Display Button to exit without saving the new
setting and to return to the home display screen.
Off
Off rE*
(Alarm
off with
reminder.)
* When Alarm On/Off is set to "Off rE", the audible alarm "beeps" twice every three minutes to remind
the user that the Rad-87 is currently in alarm status but the audible alarm is muted. Visual alarms are
active in this mode. If an alarm limit is violated, the associated parameter/measurement label and value
flash, the alarm bell flashes red for high priority alarms and the System Status Light is solid yellow for
low priority alarms, flashes yellow for medium priority alarms or flashes red for high priority alarms.