User's Manual
4-24
Rad-87 Pulse CO-Oximeter Operator’s Manual
4
Operation
Setup Menu Level 2, continued
Button Volume
BUTTON SETTING
Press the Enter Button
again to move to the
next menu.
5x
Level 2
(Default)
Use Up or Down Button to move between
settings and the Enter Button to accept the
setting and move to the next menu screen.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
Level 1
Off
Level 3
FastSat
FastSat enables rapid tracking of arterial oxygen saturation changes. Arterial oxygen saturation
data is averaged using pulse oximeter averaging algorithms to smooth the trend. When the Rad-
87 is set to FastSat “On”, the averaging algorithm evaluates all the saturation values providing
an averaged saturation value that is a better representation of the patient’s current oxygenation
status. With FastSat, the averaging time is dependent on the input signal.
BUTTON SETTING
Press the Enter Button
again to move to the
next menu.
6x
Off
(Default)
Use Up or Down Button to move
between settings and the Enter Button
to accept the setting and move to the
Alarm Volume menu.
OR
press the Display Button to exit without
saving the new setting and to return to
the home display screen.
On