User's Manual
Rad-87 Pulse CO-Oximeter Operator’s Manual
4-9
4
LED
NORM
APOD
MAX
SENSITIVITY
MODE
rbc monitor
Rad-87
APOD
APOD
MAX
MAX
SENSITIVITY
DISPLAY
NORM
NORM
ENTER
NORM
APOD
MAX
SENSITIVITY
MODE
rbc monitor
Rad-87
APOD
APOD
MAX
MAX
SENSITIVITY
DISPLAY
NORM
NORM
ENTER
SeN
Hrs
60
SIQa
The Rad-87 LED displays messages to communicate the remaining sensor time available to the
user, as defined in the following table:
Message
Number
LED
Message
Cause Condition
Message
Length
Audible
Tone
Recommendation
1*
SEN
nin
000
SpHb sensor is
non-functional
Not monitoring
patient,SpHb
sensor is non-
functional
Until new SpHb
sensor attached
No Replace SpHb Sensor
2
SEN
HrS
###
SpHb sensor is
connected after
Rad-87 is powered
on
Not monitoring
a patient
120 seconds No
Replace SpHb Sensor
SpHb sensor is
removed from the
patient
N/A
3
†
SEN
nin
###
SpHb sensor is
connected after Rad-
87 is powered on
Not monitoring
a patient
120 seconds No
Replace SpHb sensor
SpHb sensor is
removed from the
patient
N/A
4*
†
Rpl SEN
Sensor is non-
functional
Not monitoring
a patient
Until working
sensor attached
No
Attach working sensor
Ye s
Monitoring a
patient
5
INC
SEN
Incompatible sensor
Monitoring or
not monitoring
a patient
Until compatible
sensor attached
No
Replace with a
compatible Masimo
sensor (Refer to
Section 8)
SpHb sensor
attached to Rad-
87 without SpHb
installed
Until compatible
SpHb sensor
attached
Contact your local
Masimo Representative
to learn more about the
optional SpHb upgrade
OR use a compatible
non-SpHb sensor
* While actively monitoring a patient, if the sensor time on the Rainbow reusable SpHb sensor reaches 0 minutes
remaining, the sensor will NOT stop monitoring the patient until the sensor is off the patient and no pulse is detected.
† “###” represents the numerical value of either hours or minutes remaining on the attached Rainbow reusable SpHb
sensor.
Operation
Light Emitting Diodes (LED) Display