User's Manual

7-2 Radical-7R Pulse CO-Oximeter Operator’s Manual
7
Arterial Oxygen Saturation (%SpO
2
)1%
Carboxyhemoglobin Saturation (%SpCO) 1%
Methemoglobin Saturation (%SpMet) 0 .1%
Total Hemoglobin (SpHb g/dl)
0.1 g/dl
Respiration Rate (RRa) 1 breath per minute
Pulse Rate (bpm) 1 bpm
ELECTRICAL
Standalone
AC Power requirements: 100-240 VAC, 47-63 Hz
Power consumption: 55 VA
Fuses: 1 Amp, Fast Acting, Metric, (5x20mm), 250V
Batteries
Handheld:
Type: NiMH
Capacity: 4 hours
8
Charging time: 3 hours
Docking Station (RDS-1B):
Type: NiMH
Capacity: 10 hours
8
Charging time: 6 hours
ENVIRONMENTAL
Operating Temperature: 41°F to 104°F (5°C to 40°C)
Transport/Storage Temperature: -40°F to 158°F (-40°C to +70°C)
9
Operating Humidity: 5% to 95%, non-condensing
Operating Altitude:
500 mbar to 1060 mbar pressure
-1000 ft to 18,000 ft (-304 m to 5,486 m)
PHYSICAL CHARACTERISTICS
Dimensions
Handheld: 8.9” x 3.5” x 2.1” (22.6 cm x 8.9cm x 5.3 cm)
Standalone: 3.5” x 10.5” x 7.7” (8.9 cm x 26.7cm x 19.6cm)
Weight
Handheld: 1.2 lbs. (0.54 kg)
Docking Station (RDS-1, RDS-2, RDS-3): 2.5 lbs. (1.14 kg)
Docking Station (RDS-1B): 4.11 lbs (1.86 kg)
Standalone (RDS-1, RDS-2, RDS-3): 3.8 lbs. (1.73 kg)
Standalone (RDS-1B): 5.4 lbs. (2.45 kg)
Mode
Averaging mode: 2, 4, 8,10, 12, 14 or 16 seconds
10
Sensitivity: Normal and Maximum
11
and APOD
Alarms
Specifications
Radical-7R specifications (contin