User's Manual

Radical-7R Pulse CO-Oximeter Operator’s Manual 7-1
7
Specifications
Radical-7R specifications
PERFORMANCE
Measurement Range
SpO
2
: 0 -100%
SpMet: 0 - 99.9%
SpCO: 0 - 99%
SpHb 0 - 25 g/dl
SpOC 0 - 35 ml of O
2
/dl of blood
SpHct 0 - 75%
Pulse Rate: 25 - 240 (bpm)
RRa (Respiration Rate) 0-70 breaths per minute
SIQa 0-100%
Perfusion Index: 0.02% - 20%
Pleth Variability Index: 0 - 100%
ACCURACY
Arterial Oxygen Saturation Accuracy
1
Saturation 60% to 80%
No Motion
Adults, Infants, Pediatrics ±3%
Saturation 70% to 100%
No Motion
2
Adults, Infants, Pediatrics ± 2%
Neonates ± 3%
Motion
3
Adults, Infants, Pediatrics, Neonates
± 3%
Low Perfusion
4
Adults, Infants, Pediatrics, Neonates ± 2%
Pulse Rate Accuracy
5
Pulse rate:
25 - 240 bpm
No Motion
Adults, Infants, Pediatrics, Neonates
± 3 bpm
Motion
Adults, Infants, Pediatrics, Neonates
± 5 bpm
Low Perfusion
4
Adults, Infants, Pediatrics, Neonates
± 3 bpm
Carboxyhemoglobin saturation accuracy (%SpCO)
1
Adults, Infants, Pediatrics
1% - 40% ± 3%
Methemoglobin saturation accuracy (%SpMet)
1
Adults, Infants, Pediatrics, Neonates 1% - 15% ± 1%
Total Hemoglobin accuracy (SpHb g/dl)
6
Adults, Pediatrics 8 - 17 g/dl ±1 g/dL
Respiratory Rate Accuracy (RRa, breaths per minute)
7
Adults 4 to 70 ± 1 breath per minute
Resolution