User's Manual
4-16 Radical-7R Pulse CO-Oximeter Operator’s Manual
4
Operation
MENU ITEMS DESCRIPTION
PI HIGH LIMIT
The PI high alarm limit can be set anywhere between 0.04 and 19, then “---”
with a 0.01 step size between 0.04 and 0.10, a 0.10 step size between 0.10
and 1.0, and a 1.0 step size between 1.0 and 19. In the “---” (off) setting,
the PI High Alarm Limit Alarm is disabled. Factory default setting is “---” (off)
PI LOW LIMIT
The PI low alarm limit can be set as “---”, or anywhere between 0.03 to
18 with a .01 step size. In the “---” (off) setting, the PI Low Alarm Limit
Alarm is disabled. Factory default setting is “---” (off).
NOTE: The low alarm limit must be set below the high alarm setting. If the high alarm limit
is set below the low alarm limit, the low alarm limit will automatically adjust to the next
setting below the newly entered high alarm limit setting.
PVI HIGH LIMIT
The PVI high alarm limit can be set anywhere between 2 and 99, then
“---” with a 1 step size between 2 and 99. In the “---” (off) setting, the PVI
High Alarm Limit Alarm is disabled. Factory default setting is “---” (off).
PVI LOW LIMIT
The PVI low alarm limit can be set as “---”, or anywhere between 1 and
98 with a 1 step size. In the “---” (off) setting, the PVI Low Alarm Limit
Alarm is disabled. Factory default setting is “---” (off).
NOTE: The low alarm limit must be set below the high alarm setting. If the high alarm
limit is set below the low alarm limit, the low alarm limit will automatically adjust to the next
setting below the newly entered high alarm limit setting.
RRa HIGH LIMIT
The respiration rate high alarm limit can be set anywhere between
5 to 69 breaths per minute., with a 1 step size.
RRa LOW LIMIT
The respiration rate low alarm limit can be set anywhere between 4 to
68 breaths per minute, with a 1 step size.
In the “----” (off) setting, the alarm
can be turned off completely.
NOTE: The low alarm limit must be set below the high alarm setting. When the high alarm limit
is set below the low alarm limit, the low alarm limit will automatically adjust to the next setting
below the newly entered high alarm limit setting.
DELAY
This menu allows the users to set an audible saturation delay. The delay
can be set to either 0, 5,10 or 15 seconds. The delay setting only affects
saturation alarms indications.