User's Manual

Radical-7R Pulse CO-Oximeter Operator’s Manual
NON-INVASIVE TOTAL HEMOGLOBIN (SpHb) ACCURACY COMPARED TO INVASIVE
LABORATORY METHODS*
In 492 comparisons of non-invasive total hemoglobin (SpHb) and invasive hemoglobin (tHb)
measurements from a laboratory CO-Oximeter, SpHb accuracy was as follows:
0.90 correlation
0.95 g/dL standard deviation
Below 12 g/dL, 99% of SpHb readings were < 2 g/dL of the laboratory tHb value
At or above 12 g/dL, 94% of SpHb readings were < 2 g/dL of the laboratory value
* Masimo FDA Submission Data
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES
The Radical-7R Signal Extraction Pulse CO-Oximeter is designed to minimize
the possibility of hazards from errors in the software program by following sound
engineering design processes, Risk Analysis and Software Validation.
Variation in hemoglobin measurements may be profound and may be affected by
sample type, body positioning, as well as other physiological conditions. As with most
hemoglobin tests, Radical-7R test results should be scrutinized in light of a specific
patient’s condition. Any results exhibiting inconsistency with the patient’s clinical status
should be repeated and/or supplemented with additional test data.
Explosion hazard. Do not use the Pulse CO-Oximeter in the presence of
ammable anesthetics or other fl ammable substance in combination with air,
oxygen-enriched environments, or nitrous oxide.
High intensity, extreme lights (including pulsating strobe lights) directed on the sensor
may not allow the Pulse CO-Oximeter to obtain readings.
Excessive ambient noise may affect the accuracy of the respiration rate reading from the
Acoustic Respiration Sensor.
When monitoring Acoustic Respiration, Masimo recommends minimally monitoring
both oxygenation (SpO2) and respiration (RRa).
The Pulse CO-Oximeter is NOT intended for use as an apnea monitor.
Pulse rate measurement is based on the optical detection of a peripheral flow pulse
and therefore may not detect certain arrhythmias. The Pulse CO-Oximeter should
not be used as a replacement or substitute for ECG based arrhythmia analysis.
The Pulse CO-Oximeter should be considered an early warning instrument. As a
trend towards patient hypoxemia is indicated, blood samples should be analyzed
by laboratory instruments to completely understand the patient’s condition.
The Pulse CO-Oximeter is to be operated by qualifi ed personnel only. This
manual, accessory Directions for Use (DFU), all precautionary information, and
specifi cations should be read before use.
Electric shock hazard. Do not open the Pulse CO-Oximeter cover except to
replace the battery of the Handheld instrument. Only a qualifi ed operator may
perform maintenance procedures specifi cally described in this manual. Refer
servicing to Masimo for repair of this equipment.
Ensure that the HF surgical neutral electrode is properly connected to help prevent unin-
tended current return paths when using high frequency (HF) surgical equipment.
As with all medical equipment, carefully route patient cabling to reduce the
ii