Radical-7 Chapter 4: Operation
www.masimo.com 71 Masimo
In Vivo for SpCO
From the In Vivo screen, access the following screens:
Options
Description
Factory
Default
Settings
User Configurable Settings
Enabled
See Overview of In Vivo
Adjustment on page 60.
No
Yes, No
Offset Amount
See Overview of In Vivo
Adjustment on page 60.
0
Yes, No
± 9% in steps of ± 1%
SpMet Settings
From the SpMet Settings screen, access the following screens:
SpMet Alarms on page 71.
Accessing Parameter Information on page 59.
SpMet Alarms
From the Alarms screen, change any of the following options:
Options
Description
Factory
Default
Settings
User Configurable Settings
High Limit
The High Alarm Limit
is upper threshold
that triggers an
alarm.
3.0
1.0% to 2.0% in steps of 0.1
2.5% to 99.5% in steps of 0.5%, or
Off
Low Limit
The Low Alarm Limit
is lower threshold
that triggers an
alarm.
Off
Off, 0.1% to 2.0% in steps of 0.1%
2.5% to 99%, in steps of 0.5%