User's Manual

Table Of Contents
Radical-7 Chapter 4: Operation
www.masimo.com 66 Masimo
RRa Settings
From the RRa Settings screen, access any of the following screens:
RRa Alarms on page 66.
Additional Settings for RRa on page 67.
Accessing Parameter Information on page 59.
RRa Alarms
From the Alarms screen, change any of the following options:
Options
Description
Factory
Default
Settings
Configurable Options
High Limit
The High Limit is upper
threshold that triggers
an alarm.
30 breaths
per minute
5 to 69 breaths per minute in
steps of 1, or Off
Low Limit
The Low Limit is lower
threshold that triggers
an alarm.
4 breaths per
minute
4 to 68 breaths per minute in
steps of 1
Respiratory
Pause
The duration of time
that triggers an alarm if
no breaths are
detected.
15 seconds
20, 25, 30, 35, 40, or 15
seconds
Alarm Delay
When a High or Low
alarm condition occurs,
this feature delays the
audible part of an
alarm.
30 seconds
60, 0, 10, 15, or 30 seconds