Radical-7 Chapter 4: Operation
www.masimo.com 63 Masimo
SpHb Alarms
From the Alarms screen, change any of the following options:
Options
Description
Factory Default
Settings
User Configurable Settings
High Limit
The High Limit is
upper threshold
that triggers an
alarm.
17.0 g/dL
(11.0 mmol/L)
2.0 g/dL to 24.5 g/dL in steps
of 0.1 g/dL, or Off
(2.0 mmol/L to 15.0 mmol/L
in steps of 0.1 mmol/L, or Off)
When SpHb Precision is set
to 1.0, the values are rounded
down.
When set to Off, alarms are
disabled.
Low Limit
The Low Limit is
lower threshold
that triggers an
alarm.
7.0 g/dL
(4.0 mmol/L)
Off, or 1.0 g/dL to 23.5 g/dL
in steps of 0.1 g/dL
(Off, or 1.0 mmol/L to 14.5
mmol/L, in steps of 0.1
mmol/L)
When SpHb Precision is set
to 1.0, values are rounded
down.
When set to Off, alarms are
disabled.
Additional Settings for SpHb
From the Additional Settings screen, you change any of the following options:
Options
Description
Factory Default
Settings
User Configurable
Settings
Averaging Time
The length of time over
which the system
calculates the average of
all data points.
Medium
Short, Medium, or Long
Calibration
Provides an Arterial (SpHb)
or Venous (SpHb
v
) that
displays on the main
screen.
Venous
Arterial or Venous