Radical-7 Chapter 4: Operation
www.masimo.com 61 Masimo
SpO2 Settings
Access any of the following options:
Alarms for SpO2.
Additional Settings for SpO2 on page 62.
Accessing Parameter Information on page 59.
Alarms for SpO2
From the Alarms screen, change any of the following options:
Options
Description
Factory Default
Settings
Configurable Options
High Limit
High Limit is the upper
threshold that triggers an
alarm.
Off
2% to 99% in steps of
1, or Off
When set to Off,
alarms are disabled
Low Limit
Low Limit is the lower
threshold that triggers an
alarm.
88%
1% to 98% in steps of 1
Rapid Desat
Sets the Rapid Desat limit
threshold to the selected
amount below the Low
Alarm Limit. When SpO2
value falls below rapid desat
limit the audio and visual
alarm are immediately
triggered without respect to
the alarm delay.
-10%
Off, -5%, or -10%
Alarm Delay
When an alarm condition is
met, this feature delays the
audible part of an alarm.
5 seconds
0, 5, 10, or 15 seconds
Adaptive
Threshold
Alarm (ATA)
ATA establishes
patient-specific limit
thresholds based upon the
baseline value of the
parameter.
See Adaptive Threshold
Alarm (ATA) Feature on
page 97.
Off
Off or On