User's Manual

Table Of Contents
Radical-7 Chapter 3: Setup
www.masimo.com 43 Masimo
The SpO2 and pulse rate values appear on the VueLink compatible monitor.
4. In order for the plethysmographic waveform to be displayed on the VueLink
compatible monitor, and for the VueLink monitor to convey alarm conditions
measured by the Radical-7, the VueLink compatible monitor must be properly
configured.
5. See instructions for use provided with the VueLink compatible monitor and the
VueLink module. See Device Related Safety Information, Warnings, and Cautions
on page 15 and Serial Interface Specifications on page 118.
Setting Up for SpaceLabs Flexport
To set up for use with SpaceLabs Flexport
1. On the Radical-7, on the device output screen, for the serial option, select
SpaceLabs Flexport.
2. Connect one end of the Spacelabs Flexport cable to the Serial Output connector
on the Docking Station.
3. Connect the other end of the Spacelabs Flexport cable to the Spacelabs Universal
Flexport connector.
The SpO2 and pulse rate values appear on the Spacelabs screen.
4. In order for the plethysmographic waveform to be displayed on the Spacelabs
screen, and for the Spacelabs monitor to convey alarm conditions measured by
the Radical-7, the Spacelabs monitor must be properly configured.
5. See instructions for use provided with the Spacelabs monitor. See Device Related
Safety Information, Warnings, and Cautions on page 15 and Serial Interface
Specifications on page 118.