User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screenlock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Configurations
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Service Required Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Troubleshooting Measurements
- Signal Indication and Quality Indicator (SIQ)
- Dimly Lit Parameters
- Low Perfusion
- Low Signal Quality
- SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements
- Unexpected SpO2, SpCO, SpMet, or SpHb Reading
- Unexpectedly High SpCO Reading
- Difficulty Obtaining a Reading
- SpCO Reading Displays as Dashes
- Troubleshooting the Radical-7
- Troubleshooting Measurements
- Chapter 8: Specifications
- Performance
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- ISO Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Safety Information, Warnings, and Cautions
www.masimo.com 13 Masimo
Note: Inaccurate SpO2 readings may be caused by:
Elevated levels of COHb and MetHb
For increased COHb: COHb levels above normal tend to increase the level of
SpO2. The level of increase is approximately equal to the amount of COHb
that is present.
Note: High levels of COHb may occur with a seemingly normal SpO2. When
elevated levels of COHb are suspected, laboratory analysis (CO-Oximetry) of
a blood sample should be performed.
For increased MetHb: the SpO2 may be decreased by levels of MetHb of up to
approximately 10% to 15%. At higher levels of MetHb, the SpO2 may tend to read
in the low to mid 80s. When elevated levels of MetHb are suspected, laboratory
analysis (CO-Oximetry) of a blood sample should be performed.
Note: Inaccurate SpO2 readings may be caused by:
Intravascular dyes such as indocyanine green or methylene blue
Externally applied coloring and texture such as nail polish, acrylic nails, glitter,
etc.
Elevated levels of bilirubin
Severe anemia
Low arterial perfusion
Motion artifact