User's Manual

Table Of Contents
Radical-7 Chapter 7: Troubleshooting
www.masimo.com 110 Masimo
Continuous Speaker Tone
Internal failure.
Next steps: To silence an alarm, press the Alarm Silence button. If alarm continues to
sound, turn off the Radical-7. If necessary, remove Handheld battery. Requires service. See
Contacting Masimo on page 133.
Buttons Do Not Work When Pressed
Internal failure.
Next steps: Requires service. See Contacting Masimo on page 133.
Handheld Battery Does Not Charge
AC power cable may be disconnected.
Next steps: Restore power to the instrument.
Battery Run Time Significantly Reduced
Battery memory effects. Next steps: See Battery Operation and Maintenance on page 127.
Indicators on Docking Station Continuously Flash
Incompatible version of software on Handheld and Docking Station.
Next steps: Upgrade to current software versions. Match Handheld to Docking Station with
compatible software versions.