User's Manual

Table Of Contents
Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 104 Masimo
No Sensor Connected Message
Message:
(Pulse Ox) No Sensor Connected
(RAM) No Sensor Connected
Sensor not fully inserted into the connector. May be an incorrect sensor or a
defective sensor or cable.
Next steps: Disconnect and reconnect sensor. See the instructions for use
provided with your sensor.
Instrument is searching for patient’s pulse.
Next steps: Disconnect and reconnect the sensor into the Patient Cable
connector.
Sensor is disconnected from patient cable. Sensor connected upside down into
patient cable.
Next steps: Check to see if the sensor LED is flashing. Disconnect and reconnect
the sensor. If the LED fails to operate, replace the sensor.
Pulse Search Message
Message: (Pulse Ox) Pulse Search
Instrument is searching for pulse.
Next steps: If instrument fails to display within 30 seconds, disconnect and reconnect. If
pulse search continues, move sensor to better perfused site.
Sensor Initializing Message
Message: (Pulse Ox) Sensor Initializing
Instrument is checking the sensor for proper functioning and performance.
Next steps: If values are not displayed within 30 seconds, disconnect and reconnect sensor.
If values are still not displayed, replace with a new sensor.
Sensor Off Patient Message
Message:
(Pulse Ox) Sensor Off Patient
(RAM) Sensor Off Patient
Sensor off patient.
Next steps: Disconnect and reconnect sensor. Reattach sensor.
Sensor not connected to patient properly. Sensor is damaged.
Next steps: Properly reapply the sensor on the patient and reconnect the sensor to the
instrument or patient cable. If the sensor is damaged, replace the sensor.