User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screenlock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Configurations
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Service Required Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Troubleshooting Measurements
- Signal Indication and Quality Indicator (SIQ)
- Dimly Lit Parameters
- Low Perfusion
- Low Signal Quality
- SpO2 Values Do Not Correlate With Clinical Assessment or Arterial Blood Gas Measurements
- Unexpected SpO2, SpCO, SpMet, or SpHb Reading
- Unexpectedly High SpCO Reading
- Difficulty Obtaining a Reading
- SpCO Reading Displays as Dashes
- Troubleshooting the Radical-7
- Troubleshooting Measurements
- Chapter 8: Specifications
- Performance
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- ISO Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 102 Masimo
No Adhesive Sensor Connected Message
Message:
(Pulse Ox) No Adhesive Sensor Connected
(RAM) No Adhesive Sensor Connected
When a single-patient-use sensor is used, the adhesive portion of the sensor is not
connected. (Applies to Rainbow ReSposable CO-Oximeter Sensor Systems or ReSposable
Pulse CO-Oximeter Sensor Systems only.)
Next steps: Ensure the adhesive portion is firmly connected to the sensor.
Interference Detected Message
Message:
(Pulse Ox) Interference Detected
(RAM) Interference Detected
High intensity light such as pulsating strobe lights, excessive ambient light sources such as
surgical lights or direct sunlight, or other monitor displays.
Next steps: Place a Masimo Optical Light Shield over the sensor.
Incorrect monitor line frequency setting (Hz).
Next steps: . Adjust the Line Frequency to the correct Hz setting. See Device Settings on
page 74.
SpO2 Only Mode Message
Message: (Pulse Ox) SpO2 Only Mode
Occurs during an unsuccessful sensor initialization/pulse search routine or during
monitoring.
Next steps: See the directions for use provided with your sensor. Use a Masimo light shield
to cover the sensor and adjust the sensor.
Low Battery Message
Battery charge is low.
Next steps: Charge battery by placing the Handheld into the Docking Station and powering
the instrument with AC line power. Replace battery if necessary.
Low Perfusion Index Message
Message: (Pulse Ox) Low Perfusion Index
Signal too small.
Next steps: Move sensor to better perfused site. See Low Perfusion on page 108.