Radical-7 Chapter 4: Operation
www.masimo.com 95 Masimo
Options Description Factory
Default
Settings
User Configurable
Settings
High
Limit
The High Limit is
upper threshold that
triggers an alarm.
17.0 g/dL
(170 g/L)
(11.0
mmol/L)
2.0 g/dL to 24.5 g/dL in
steps of 0.1 g/dL, or Off
(2 g/L to 249 g/L in
steps of 1 g/L, or Off)
(2.0 mmol/L to 15.0
mmol/L in steps of 0.1
mmol/L, or Off)
When SpHb Precision is
set to 1.0, the values are
rounded down.
When set to Off, alarm is
disabled.