Radical-7 Chapter 4: Operation
www.masimo.com 93 Masimo
Options Description Factory
Default
Settings
User Configurable
Settings
Averaging
Time
The length of time over
which the system
calculates the average of
all data points.
8 seconds 2-4, 4-6, 8, 10,
12, 14, or 16
seconds
FastSat
See
FastSat Overview
on
page 93.
Off On or Off
FastSat Overview
FastSat enables rapid tracking of arterial oxygen saturation changes.
Arterial oxygen saturation data is averaged using pulse oximeter
averaging algorithms to smooth the trend.
When the Radical-7 is set to FastSat
On
, the averaging algorithm
evaluates all the saturation values providing an averaged saturation value
that is a better representation of the patient’s current oxygenation status.
With FastSat, the averaging time is dependent on the input signal.