Options Description
Factory
Default
Settings
Configurable
Options
Rapid
Desat
Sets the Rapid Desat limit
threshold to the selected
amount below the Low Alarm
Limit. When SpO
2
value falls
below rapid desat limit the
audio and visual alarm are
immediately triggered without
respect to the alarm delay.
-10%
Off, -5%, or -
10%
Alarm
Delay
When an alarm condition is
met, this feature delays the
audible part of an alarm.
5
seconds
0, 5, 10, or 15
seconds
Adaptive
Threshold
Alarm
(ATA)
ATA establishes patient-specific
limit thresholds based upon the
baseline value of the
parameter.
See
Adaptive Threshold alarm
(ATA) Feature
on page 156.
Off Off or On
Additional Settings for SpO2
From the
Additional Settings
screen, change any of the following options: