Operator's Manual

Table Of Contents
Radical-7 Chapter 4: Operation
www.masimo.com 91 Masimo
SpO2 Settings
Access any of the following options:
Alarms for SpO2
on page 91.
Additional Settings for SpO2
on page 92.
About Parameter Information
on page 88.
Alarms for SpO2
From the
Alarms
screen, change any of the following options:
Options Description
Factory
Default
Settings
Configurable
Options
High Limit
The High Limit is upper
threshold that triggers an
alarm.
Off
2% to 99% in
steps of 1%,
or Off
When set to
Off, alarm is
disabled
Low Limit
Low Limit is the lower threshold
that triggers an alarm.
88%
1% to 98% in
steps of 1%