Radical-7 Chapter 4: Operation
www.masimo.com 89 Masimo
2. An
About
screen appears for the selected parameter.
In Vivo Adjustment Overview
The In Vivo Adjustment feature lets clinicians manually adjust one or
more clinical parameters to match that of a corresponding laboratory
reference for continuous trending. To remind clinicians that the feature is
active, an offset value displays alongside the adjusted parameter value.
The In Vivo Adjustment feature for a parameter can be turned on by
accessing the In Vivo screen in the settings menu of that parameter. After
enabling the feature, set an offset value. Once the feature is enabled, a
positive or a negative offset value appears, as shown in the following
illustration.
The In Vivo offset is set to zero for any of the following:
Cable or sensor is disconnected from device.
Sensor goes off patient causing a sensor initialization to occur.
Eight hours has elapsed since the In Vivo value was activated.
Restore of factory defaults.
The user turns off In Vivo.