7. While in the SatShare mode, if there are any significant
discrepancies between the readings from Radical-7 and those on
the monitor displaying the values obtained from SatShare, the
values reported by the Radical-7 are to be considered the correct
values.
8. It is possible to use the Radical-7 with SatShare while the
Radical-7 is not connected to AC power. However, in this
configuration, battery run time is reduced. See
Battery
Operation and Maintenance
on page 196.
9. On the Radical-7, turn on the
Satshare Numbers
option. See
Device Output
on page 126.
10. If displaying the simulated waveform is not desirable, it is
recommended to turn off the plethysmographic waveform
display of the multi-parameter patient monitor. See
Serial
Interface Specifications
on page 186.