Operator's Manual
Table Of Contents
- About this Manual
- Product Description
- Indications for Use
- Contraindications
- Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology®
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Oxygen Reserve Index (ORI)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Standby and Power Off
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- RAM Check Sensor Message
- RAM Sensor Initializing Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- Invalid Parameter Alarm Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Near Expiration Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Chapter 9: Service and Maintenance
- Appendix
- Index
Radical-7 Chapter 3: Setup
www.masimo.com 63 Masimo
plethysmographic waveform and the parameter values. Refer to the
instructions for use provided with the multi-parameter monitor. See
Compliance
on page 183.
To set up for use with SatShare interface
1. Select the SatShare cable that is appropriate for the multi-
parameter monitor. For the latest list of available SatShare cables
and validated devices, see
www.masimo.com
.
2. Connect the labeled end of the SatShare cable to the SatShare
Cable connector on the Docking Station. See
Standalone Back
Panel
on page 56. For a secure connection, tighten the cable
connector screws.
3. Connect the other end of the SatShare cable to one of the
following:
• Sensor connector of the multi-parameter monitor cable
• Directly to the multi-parameter monitor
4. Verify that the multi-parameter monitor recognizes the SatShare
cable.
5. As appropriate, configure alarm limits on the multi-parameter
monitor.
6. Set the averaging time for the multi-parameter monitor to its
lowest setting (or fastest response). The ideal waveform for the
Radical-7 requires additional averaging by the monitor. If the
averaging time of the multi-parameter monitor is not changed,
the time to display physiological changes in saturation on the
monitor is increased with SatShare. However, the delay can be
minimized by reducing the averaging time on the multi-
parameter monitor.