Radical-7 Chapter 3: Setup
www.masimo.com 63 Masimo
plethysmographic waveform and the parameter values. Refer to the
instructions for use provided with the multi-parameter monitor. See
Compliance
on page 183.
To set up for use with SatShare interface
1. Select the SatShare cable that is appropriate for the multi-
parameter monitor. For the latest list of available SatShare cables
and validated devices, see
www.masimo.com
.
2. Connect the labeled end of the SatShare cable to the SatShare
Cable connector on the Docking Station. See
Standalone Back
Panel
on page 56. For a secure connection, tighten the cable
connector screws.
3. Connect the other end of the SatShare cable to one of the
following:
Sensor connector of the multi-parameter monitor cable
Directly to the multi-parameter monitor
4. Verify that the multi-parameter monitor recognizes the SatShare
cable.
5. As appropriate, configure alarm limits on the multi-parameter
monitor.
6. Set the averaging time for the multi-parameter monitor to its
lowest setting (or fastest response). The ideal waveform for the
Radical-7 requires additional averaging by the monitor. If the
averaging time of the multi-parameter monitor is not changed,
the time to display physiological changes in saturation on the
monitor is increased with SatShare. However, the delay can be
minimized by reducing the averaging time on the multi-
parameter monitor.