Radical-7 Chapter 3: Setup
www.masimo.com 62 Masimo
Setting Up for SpaceLabs Flexport
To set up for use with SpaceLabs Flexport
1. On the Radical-7, on the
device output
screen, for the
serial
option, select SpaceLabs Flexport.
2. Connect one end of the Spacelabs Flexport cable to the Serial
Output connector on the Docking Station.
3. Connect the other end of the Spacelabs Flexport cable to the
Spacelabs Universal Flexport connector.
The SpO
2
and pulse rate values appear on the Spacelabs screen.
4. In order for the plethysmographic waveform to be displayed on
the Spacelabs screen, and for the Spacelabs monitor to convey
alarm conditions measured by the Radical-7, the Spacelabs
monitor must be properly configured.
5. See instructions for use provided with the Spacelabs monitor.
See
Compliance
on page 183 and
Serial Interface
Specifications
on page 186.
Setting Up and Using SatShare
Parameter values from the Radical-7 can be displayed on a multi-
parameter monitor through the SatShare feature. The SatShare feature
provides an ideal, simulated plethysmographic waveform that
corresponds to the parameter values determined the by Radical-7. This
waveform may be used to display these values on multi-parameter
monitors through the multi-parameter oximetry sensor or input
connector.
It is recommended that the Radical-7 be positioned near the multi-
parameter monitor, with the Radical-7 screen displaying the