Operator's Manual
Table Of Contents
- About this Manual
- Product Description
- Indications for Use
- Contraindications
- Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology®
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Oxygen Reserve Index (ORI)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Standby and Power Off
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- RAM Check Sensor Message
- RAM Sensor Initializing Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- Invalid Parameter Alarm Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Near Expiration Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Chapter 9: Service and Maintenance
- Appendix
- Index
www.masimo.com 59 Masimo
Chapter 3: Setup
The following chapter contains information about setting up the Radical-7
before use.
Unpacking and Inspection
To unpack and inspect the device
1. Remove the device from the shipping carton and examine it for
signs of shipping damage.
2. Check all materials against the packing list. Save all packing
materials, invoice and bill of lading. These may be required to
process a claim with the carrier.
3. If anything is missing or damaged, contact the Technical Service
Department. See
Return Procedure
on page 204.
Docking Station Power Requirements
• Always use a hospital-grade, AC power cable to connect the
Docking Station to an AC power source.
• Do not connect the Docking Station to an AC outlet that is
controlled by a switch because the power to the Docking Station
may be inadvertently switched off.
• Verify the AC power voltage and line frequency before use.
• Verify that the power source can provide an adequate power
rating as indicated on the rear panel of the Docking Station.
• The Radical-7 is designed to operate on 100 to 240VAC, 47-63
Hz.
• The Radical-7 is rated at 55 VA max.