Operator's Manual

Table Of Contents
Radical-7 Chapter 2: Radical-7 Descriptions
www.masimo.com 57 Masimo
Item and Description
2. Analog Output/Nurse Call connector
Use the Analog Output connector with a ferrite bead installed to
interface with an analog output device, such as a chart recorder or
Nurse Call system. All external device connections to the Analog
Output/Nurse Call connector must be IEC-60950 compliant.
See
Serial Interface Specifications
on page 186.
3. SatShare Cable connector
Use the SatShare Cable connector to connect a SatShare cable to the
SpO
2
input connector of a multi-parameter patient monitor. All external
device connections to the SatShare Cable Connector must be IEC-
60601-1-1 compliant. SatShare cables are available to interface with
most major multi-parameter patient monitors. Check the label on the
SatShare cable and the SatShare Directions for Use to ensure that the
correct cable is used for each type of patient monitor.
Visit
www.masimo.com
for the latest SatShare cables and validated
devices.
4. Power Entry module
The Power Entry module contains the input connector for AC power
and two fuses. The AC input provides power to the system from the AC
line. Always connect the Radical-7 to the mains power for continuous
operation and/ or battery recharging. Note: Use the power cord as the
means to disconnect the device from the mains power supply.