Operator's Manual
Table Of Contents
- About this Manual
- Product Description
- Indications for Use
- Contraindications
- Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology®
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Oxygen Reserve Index (ORI)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Standby and Power Off
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- RAM Check Sensor Message
- RAM Sensor Initializing Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- Invalid Parameter Alarm Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Near Expiration Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Chapter 9: Service and Maintenance
- Appendix
- Index
Radical-7 Warnings and Cautions
www.masimo.com 26 Masimo
Warning: Do not use petroleum-based or acetone solutions, or other
harsh solvents, to clean the Radical-7. These substances affect the
device’s materials and device failure can result.
Warning: Do not touch, press, or rub the display panels with abrasive
cleaning compounds, instruments, brushes, rough-surface materials, or
bring them into contact with anything that could scratch the display.
Caution: An operator may only perform maintenance procedures
specifically described in the manual. Refer servicing to qualified service
personnel trained in the repair of this equipment.
Caution: Do not submerge Radical-7 in any cleaning solution or attempt
to sterilize by autoclave, irradiation, steam, gas, ethylene oxide or any
other method. This will seriously damage Radical-7.
Note: Changes or modifications not expressly approved by Masimo shall
void the warranty for this equipment.
Note: Excessive cleaning solution can flow into the monitor and cause
damage to internal components.
Warning: Fire Hazard: To protect against fire hazard, replace only with
fuses of same type, current rating, and voltage rating.
Warning: Electrical Shock Hazard: The Docking Station battery should be
installed and/or removed from the Docking Station only by qualified
personnel.
Caution: Electrical Shock Hazard: Carry out periodic tests to verify that
leakage currents of patient-applied circuits and the system are within
acceptable limits as specified by the applicable safety standards. The
summation of leakage currents must be checked and in compliance with
IEC 60601-1 and UL60601-1. The system leakage current must be checked
when connecting external equipment to the system. When an event such