Operator's Manual
Table Of Contents
- About this Manual
- Product Description
- Indications for Use
- Contraindications
- Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology®
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Oxygen Reserve Index (ORI)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Standby and Power Off
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- RAM Check Sensor Message
- RAM Sensor Initializing Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- Invalid Parameter Alarm Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Near Expiration Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Chapter 9: Service and Maintenance
- Appendix
- Index
Radical-7 Chapter 9: Service and Maintenance
www.masimo.com 205 Masimo
7. Repeat steps 5 and 6, with the positive lead of the voltmeter
connected to pin 15 (
analog 2
). See
Serial Interface
Specifications
on page 186.
8. Connect a patient cable and sensor and verify that the voltage
on pins 9 and 15 are between 0V and 1.0V while measuring a
saturation and pulse rate.
Battery Test
To conduct a Battery test
1. Fully charge the Radical-7 by placing the Handheld into the
Docking Station and then connect the AC power.
2. Verify that the Handheld Battery Charging indicator is
illuminated.
3. When the Radical-7 is fully charged, the Handheld Battery
Charging indicator turns off.
4. Turn on the Radical-7 on and verify that the Battery indicator
shows a full charge.
Repair Policy
Masimo or an authorized service department must perform warranty
repair and service. Do not use malfunctioning equipment. Have the device
repaired.
Clean contaminated and/or dirty equipment before returning, following
the cleaning procedure described in Cleaning. Make sure the equipment
is fully dry before packing.
To return the device for service, refer to
Return Procedure
on page 204.