7. Repeat steps 5 and 6, with the positive lead of the voltmeter
connected to pin 15 (
analog 2
). See
Serial Interface
Specifications
on page 186.
8. Connect a patient cable and sensor and verify that the voltage
on pins 9 and 15 are between 0V and 1.0V while measuring a
saturation and pulse rate.
Battery Test
To conduct a Battery test
1. Fully charge the Radical-7 by placing the Handheld into the
Docking Station and then connect the AC power.
2. Verify that the Handheld Battery Charging indicator is
illuminated.
3. When the Radical-7 is fully charged, the Handheld Battery
Charging indicator turns off.
4. Turn on the Radical-7 on and verify that the Battery indicator
shows a full charge.
Repair Policy
Masimo or an authorized service department must perform warranty
repair and service. Do not use malfunctioning equipment. Have the device
repaired.
Clean contaminated and/or dirty equipment before returning, following
the cleaning procedure described in Cleaning. Make sure the equipment
is fully dry before packing.
To return the device for service, refer to
Return Procedure
on page 204.