Radical-7 Chapter 9: Service and Maintenance
www.masimo.com 202 Masimo
The device is ready to be reconnected to AC power. If the fuses
blow shortly after replacement, the device requires service.
Performance Verification
Under normal operation, no internal adjustment or recalibration is
required. Safety tests and internal adjustments should be done by
qualified personnel only. Safety checks should be performed at regular
intervals or in accordance with local and governmental regulations.
To test the performance of the Radical-7 following repairs or during
routine maintenance, follow the procedure outlined in this chapter. If the
Radical-7 fails any of the described tests, discontinue its use and correct
the problem before returning the device back to the user.
Before performing the following tests, do the following:
Place the Handheld into the Docking Station.
Connect the Docking Station to AC power and fully charge the
Handheld battery.
Disconnect any patient cables or pulse oximetry probes.
Disconnect any SatShare, serial or analog output cables from the
device.
Set the Radical-7 to Normal operating mode by going to the
Main menu and the Home Use feature to
No
.
Power-On Self Test
To conduct a Power-On Self Test
1. Connect the Battery Module to the Device Module. Refer to
Setup for instructions on how to connect the Battery Module to
the Device Module.