Operator's Manual
Table Of Contents
- About this Manual
- Product Description
- Indications for Use
- Contraindications
- Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology®
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Oxygen Reserve Index (ORI)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Standby and Power Off
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- RAM Check Sensor Message
- RAM Sensor Initializing Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- Invalid Parameter Alarm Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Near Expiration Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Chapter 9: Service and Maintenance
- Appendix
- Index
Radical-7 Chapter 9: Service and Maintenance
www.masimo.com 198 Masimo
Battery Operation and Maintenance
The Radical-7 Handheld includes a lithium ion rechargeable battery. The
Radical-7 Docking Station may include the optional 6.5 amp-hour nickel
metal hydride rechargeable battery.
Before using the Radical-7 as a Handheld or as a transport monitor, the
Handheld rechargeable battery and the optional Docking Station
rechargeable battery must be fully charged.
To charge the Handheld rechargeable battery and the Docking Station
rechargeable battery
1. Attach the Handheld to the Docking Station.
2. Connect the Docking Station to AC power.
3. Verify that the batteries are charging.
• The Docking Station Battery Charging indicator momentarily
flashes and then remains illuminated while the batteries are
actively charging.
• If the internal battery temperature exceeds recommended
operating conditions for proper battery charging, the
Handheld Battery Charging indicator continuously flashes.
When the temperature returns to recommended operating
conditions, proper battery charging proceeds.
• The Handheld battery requires approximately 4 hours for
charging. The Docking Station battery requires
approximately 6 hours for charging.
• When both the Handheld Battery Charging indicator and the
Docking Station Battery Charging indicator turn off,
additional trickle charging may occur to complete charging.