Radical-7 Chapter 8: Specifications
www.masimo.com 191 Masimo
Pin
Signal Name
10 Ground
11 Ground
12 Nurse Call -Common
13 Ground
14 Ground
15 Analog 2
Analog Output
The Radical-7® Pulse CO-Oximeter® can interface with various analog
recording devices or strip chart recorders through its Analog Output
connector located on the back of the Docking Station. The output signals
vary from approximately 0 to 1 volt in a linear fashion. The actual analog
output voltage generated may not exactly range between 0.0V to 1.0V. A
variance of ± 40 mV is acceptable.
Calibration
For device calibration purposes, the analog output signals can be set to
either 0 Volts or 1 Volt. Calibrate the analog recording system to those
levels before use.
Nurse Call
The Nurse Call feature is available when the Radical-7 is operating as a
standalone. Nurse Call is based on the relay closing or opening
depending on alarm, Low Signal IQ events, or both. For maximum