Operator's Manual

Table Of Contents
Radical-7 Chapter 7: Troubleshooting
www.masimo.com 176 Masimo
the condition persists, requires service. See.
Contacting Masimo
on page
204.
Continuous Speaker Tone
Internal failure.
Next steps: To silence an alarm, press the
Alarm Silence
button. If alarm
continues to sound, turn off the Radical-7. If necessary, remove Handheld
battery. Requires service. See
Contacting Masimo
on page 204.
Buttons Do Not Work When Pressed
Internal failure.
Next steps: Requires service. See
Contacting Masimo
on page 204.
Handheld Battery Does Not Charge
AC power cable may be disconnected.
Next steps: Restore power to the device.
Battery Run Time Significantly Reduced
Battery memory effects. Next steps: See Battery Operation and
Maintenance on page 196.
Indicators on Docking Station Continuously Flash
Incompatible version of software on Handheld and Docking Station.
Next steps: Upgrade to current software versions. Match Handheld to
Docking Station with compatible software versions.