Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 169 Masimo
Sensor not connected to patient properly. Sensor is damaged.
Next steps: Properly reapply the sensor on the patient and reconnect the
sensor to the device or patient cable. If the sensor is damaged, replace
the sensor.
Incompatible Cable Message
Message:
(Pulse CO-Ox) Incompatible Cable
Not a proper cable.
Next steps: Replace with a proper cable.
Near Expiration Message
Messages:
(Pulse CO-Ox) Cable Near Expiration
(RAM) Cable Near Expiration
Patient cable has less than 10% of active monitoring life remaining.
Next steps: Replace with new patient cable.
Messages:
(Pulse CO-Ox) Sensor Near Expiration
(RAM) Sensor Near Expiration
Reusable sensor has less than 10% active monitoring life remaining.
Next steps: Replace with new reusable sensor.
Messages:
(Pulse CO-Ox) Adhesive Near Expiration