Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 168 Masimo
Sensor is disconnected from patient cable. Sensor connected upside
down into patient cable.
Next steps: Check to see if the sensor LED is flashing. Disconnect and
reconnect the sensor. If the LED fails to operate, replace the sensor.
Pulse Search Message
Message: (
Pulse CO-Ox
)
Pulse Search
Device is searching for pulse.
Next steps: If device fails to display within 30 seconds, disconnect and
reconnect. If pulse search continues, move sensor to better perfused site.
Sensor Initializing Message
Message:
(Pulse CO-Ox) Sensor Initializing
Device is checking the sensor for proper functioning and performance.
Next steps: If values are not displayed within 30 seconds, disconnect and
reconnect sensor. If values are still not displayed, replace with a new
sensor.
Sensor Off Patient Message
Message:
(Pulse CO-Ox) Sensor Off Patient
, or
(RAM) Sensor Off Patient
Sensor off patient.
Next steps: Disconnect and reconnect sensor. Reattach sensor.