Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 167 Masimo
Invalid Parameter Alarm Message
Denoted by dashes shown as parameter value.
Unable to provide a SpO2 or PR parameter value.
Next steps: Check patient's vital condition.
No Cable Connected Message
Message:
(Pulse CO-Ox) No Cable Connected
, or
(RAM) No Cable Connected
Cable not attached or not fully inserted into the connector.
Next steps: Disconnect and reconnect cable into connector.
No Sensor Connected Message
Message:
(Pulse CO-Ox) No Sensor Connected
, or
(RAM) No Sensor Connected
Sensor not fully inserted into the connector. May be an incorrect sensor
or a defective sensor or cable.
Next steps: Disconnect and reconnect sensor. See the instructions for use
provided with your sensor.
Device is searching for patient’s pulse.
Next steps: Disconnect and reconnect the sensor into the Patient Cable
connector.