Operator's Manual
Table Of Contents
- About this Manual
- Product Description
- Indications for Use
- Contraindications
- Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology®
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Oxygen Reserve Index (ORI)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Standby and Power Off
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- RAM Check Sensor Message
- RAM Sensor Initializing Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- Invalid Parameter Alarm Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Near Expiration Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Chapter 9: Service and Maintenance
- Appendix
- Index
Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 165 Masimo
RAM unable to collect pulsing through RAM Sensor.
Next steps: Ensure proper sensor application. Check that no object is
pulling on the sensor cable causing the sensor to peel off.
RAM Sensor Initializing Message
Message:
(RAM) Sensor Initializing
Device is checking the sensor for proper functioning and performance.
Next steps: If values are not displayed within 30 seconds, disconnect and
reconnect sensor. If values are still not displayed, replace with a new
sensor.
Low Battery Message
Battery charge is low.
Next steps: Charge battery by placing the Handheld into the Docking
Station and powering the device with AC line power. Replace battery if
necessary.
Low Perfusion Index Message
Message:
(Pulse CO-Ox) Low Perfusion Index
Signal too small.
Next steps: Move sensor to better perfused site. See
Low Perfusion
on
page
173.
Low Signal IQ Message
Message:
(Pulse CO-Ox) Low Signal IQ