Operator's Manual
Table Of Contents
- About this Manual
- Product Description
- Indications for Use
- Contraindications
- Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology®
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Oxygen Reserve Index (ORI)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Standby and Power Off
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- RAM Check Sensor Message
- RAM Sensor Initializing Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- Invalid Parameter Alarm Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Near Expiration Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Chapter 9: Service and Maintenance
- Appendix
- Index
Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 157 Masimo
it also serves as absolute low alarm limit protection. SpO
2
values
exceeding the Rapid Desat low alarm limit, whether rapid or not, will
activate an audible alarm. The user can change the Rapid Desat default
setting from 5% to 10%. ATA does not allow a Rapid Desat default
setting of 0%.
When ATA is turned
Off
, the device uses the standard alarm limits and
standard alarm delays.
About Alarms
on page 153.
Alarms for SpO2
on page 91.
3D Alarms
3D Alarms include the Desat Index Alarm and the PI Delta Alarm.
Desat Index Alarm Overview
on page 157.
Perfusion Index (PI) Delta Alarm Overview
on page 159.
Desat Index Alarm Overview
The 3D Desat Index Alarm is a user-selectable feature which allows a
clinician to request an audible and visual alarm if a patient experiences a
specified number of desaturations over a specific period of time.
Traditional high and low SpO
2
alarm limits alert clinicians to saturation
levels that exceed user selected thresholds, and these thresholds are