Operator's Manual
Table Of Contents
- About this Manual
- Product Description
- Indications for Use
- Contraindications
- Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology®
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Oxygen Reserve Index (ORI)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Standby and Power Off
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- RAM Check Sensor Message
- RAM Sensor Initializing Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- Invalid Parameter Alarm Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Near Expiration Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Chapter 9: Service and Maintenance
- Appendix
- Index
Radical-7 Chapter 5: Profiles
www.masimo.com 150 Masimo
Silence
Duration
The amount of time for
which the audible part of an
alarm will be silenced. See
Silencing the Alarms
on
page
153.
2 min 1 min, 2 min, or
3 min
Smart
Tone
Allows the audible pulse to
continue to beep when the
pleth graph shows signs of
motion.
Off On or Off
Replacing Factory Default Settings for Adult and Neo Profiles
The
Adult
profile and the
Neonatal
profile can be modified to meet
specific requirements and then they can replace the factory default
settings for
Adult
and
Neonatal
profiles. As such, after a power cycle, the
Radical-7 remembers the preferred settings for
Adult
and
Neonatal
profiles instead of the factory default settings. When preferred settings
for
Adult
and
Neonatal
are saved instead of the factory default settings,
the Profile button changes to the same blue or pink color respectively.
See
Profiles Overview
on page 145.
A user can also load preferred profile configurations into the Radical-7
using a separate tool.
To change the factory default settings for Adult or Neonatal profile settings
1. Make the preferred changes to any of the Radical-7 settings.
2. Navigate to the
Access Control
screen. See
Access Control
on
page 117.