Chapter 5: Profiles
The Radical-7 can be configured for various patient types.
Profiles Overview
The Radical-7 contains a
Profiles
screen, which lets the user customize
different settings for different patient populations:
Adult
Adult profile is the factory default profile. Displays in the Status
bar as
ADULT
and the color of the Profile button turns blue.
Pediatric
Displays in the Status bar as
PEDIATRIC
and the color of the
Profile button turns pink.
Neonatal
Displays in the Status bar as
NEO
and the color of the Profile
button turns pink.
Custom
Displays in the Status bar as
CUSTOM
and the Profile button is
not illuminated and appears gray.
If no changes are made to settings, then after a power cycle, the Radical-
7 automatically resets to the
Adult
profile because
Adult
is the factory
default profile.
If the Profile setting is changed to
NEO
or
CUSTOM
, then after a power
cycle, the Radical-7 remembers the previously selected Profile setting.