Radical-7 Chapter 4: Operation
www.masimo.com 101 Masimo
Options Description Factory
Default
Settings
User
Configurable
Settings
Averaging
Time
The length of time over
which the system calculates
the average of all data
points.
Long Short or Long
Respiration Rate (RR)
The Radical-7 can determine respiration rate (RR) either by the acoustic
signal (RRa) or by the plethysmographic waveform (RRp).
RRp Settings
When using a pulse oximetry or pulse CO-Oximetry sensor with the
Radical-7, respiration rate can be determined by the plethysmographic
waveform (RRp). This method measures a patient's respiratory rate based
on plethysmographic amplitude changes that correspond to the
respiratory cycle. When using a pulse oximetry or pulse CO-Oximetry
sensor, RRp alarms and RRp settings are active and the
Display View
conveys respiratory rate as
RRp
, as shown below.