User's Manual

Table Of Contents
Radical-7 Chapter 5: Profiles
www.masimo.com 97 Masimo
Silence
Duration
The amount of time for which the
audible part of an alarm will be
silenced. See Silencing the
Alarms on page 99.
120 30, 60, 90, 120
seconds
Smart Tone Allows the audible pulse to
continue to beep when the pleth
graph shows signs of motion.
Off On or Off
Replacing Factory Default Settings for Adult and Neo Profiles
The Adult profile and the Neonatal profile can be modified to meet specific requirements and
then they can replace the factory default settings for Adult and Neonatal profiles. As such,
after a power cycle, the Radical-7 remembers the preferred settings for Adult and Neonatal
instead of the factory default settings. When preferred settings for Adult and Neonatal are
saved instead of the factory default settings, the Profile button changes to same blue or pink
color. See Profiles Overview on page 93.
A user can also load preferred profile configurations into the Radical-7 using a separate tool.
To change the factory default settings for Adult or Neonatal profile settings
1. Make the preferred changes to any of the Radical-7 settings.
2. Navigate to the Access Control screen. See Access Control on page 76.
3. For either Adult or Neonatal, touch Save.
4. Touch Ok.
5. Alternatively, the user can restore all Profile settings to their factory default values
by touching Restore, and then touching Ok.
6. Confirm the changes by powering off and powering on the Radical-7 and then
verifying settings.