User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 4: Operation
www.masimo.com 77 Masimo
Access Control Screen
From the Access Control screen, change any of the following options:
Options Description Factory
Default
Settings
User Configurable Settings
All Mute All patient alarm
conditions are
silenced. Only
system alarms will
be indicated by an
audible alarm.
Disabled Enabled or disabled
If enabled, All Mute and All Mute
with Reminder become available
settings from the Silence Duration
option on the Sounds screen. See
Sounds on page 74.
Lock Alarm
Volume
When set to 3 or 4, 3
or 4 shows dimly lit
in the Alarm Volume
section of the
Alarms Menu screen
and cannot be
changed.
Off 3, 4, or Off
SpO2 Low %
Limit
Threshold at which
SpO2 Low Alarm
Limit cannot be
reduced.
Off 1% to 98% in steps of 1, or Off
Lock Layout Prevents the user
from making
changes to the
parameter layout.
N/A On or Off
Screen Lock Prevents
unintentional
interaction with
Display View.
On On or Off
Legacy Mode Changes the Display
View from color to
monochrome.
Color Mono or Color
Save as Adult Saves
pre-configured
profiles for adult
patients.
N/A Press Save to load all device
configuration settings to adult
profile.
Save as Neo Saves
pre-configured
profiles for neonatal
patients
N/A Press Save to load all device
configuration settings to neonatal
profile.