User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 4: Operation
www.masimo.com 72 Masimo
In Vivo for SpCO
From the In Vivo screen, access the following screens:
Options Description Factory
Default
Settings
User Configurable Settings
Enabled See In Vivo Adjustment
Overview on page 61.
Off On or Off
Offset Amount See In Vivo Adjustment
Overview on page 61.
0 ± 9% in steps of 0.1%
SpMet Settings
From the SpMet Settings screen, access the following screens:
SpMet Alarms on page 72.
About Parameter Information on page 60.
SpMet Alarms
From the Alarms screen, change any of the following options:
Options Description Factory
Default
Settings
User Configurable Settings
High Limit The High Alarm Limit
is upper threshold
that triggers an alarm.
3.0 1% to 2% in steps of 0.1%
2.5% to 99.5% in steps of 0.5%, or
Off
Low Limit The Low Alarm Limit
is lower threshold that
triggers an alarm.
Off Off, 0.1% to 2.0% in steps of 0.1%
2.5% to 99%, in steps of 0.5%