Radical-7 Chapter 4: Operation
www.masimo.com 72 Masimo
In Vivo for SpCO
From the In Vivo screen, access the following screens:
Options Description Factory
Default
Settings
User Configurable Settings
Enabled See In Vivo Adjustment
Overview on page 61.
Off On or Off
Offset Amount See In Vivo Adjustment
Overview on page 61.
0 ± 9% in steps of 0.1%
SpMet Settings
From the SpMet Settings screen, access the following screens:
SpMet Alarms on page 72.
About Parameter Information on page 60.
SpMet Alarms
From the Alarms screen, change any of the following options:
Options Description Factory
Default
Settings
User Configurable Settings
High Limit The High Alarm Limit
is upper threshold
that triggers an alarm.
3.0 1% to 2% in steps of 0.1%
2.5% to 99.5% in steps of 0.5%, or
Off
Low Limit The Low Alarm Limit
is lower threshold that
triggers an alarm.
Off Off, 0.1% to 2.0% in steps of 0.1%
2.5% to 99%, in steps of 0.5%