User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 4: Operation
www.masimo.com 68 Masimo
Options Description Factory Default Configurable Options
High Limit The High Limit is
upper threshold that
triggers an alarm.
30 breaths per
minute
6 breaths per minute to 69
breaths per minute, or Off
Low Limit The Low Limit is lower
threshold that
triggers an alarm.
6 breaths per
minute
5 breaths per minute to 68
breaths per minute, or Off
Alarm Delay When an alarm
condition is met, this
feature delays the
audible part of an
alarm.
30 second 0, 10, 15, 30, 60 seconds
Additional Settings for RRp
From the Additional Settings screen, change any of the following options:
Options Description Factory Default
Settings
User Configurable Settings
Averaging
Time
The length of time
over which the system
calculates the average
of all data points.
Slow No, Fast, Medium, Slow,
Trending
Freshness The duration of time
that, during
interference, the
system displays the
last valid reading.
5 minutes 0, 1, 5, 10, 15 minutes