User's Manual

Table Of Contents
Radical-7 Chapter 4: Operation
www.masimo.com 64 Masimo
SpHb Alarms
From the Alarms screen, change any of the following options:
Options Description Factory Default
Settings
User Configurable Settings
High Limit The High Limit is
upper threshold
that triggers an
alarm.
17.0 g/dL
(11.0 mmol/L)
2.0 g/dL to 24.5 g/dL in steps
of 0.1 g/dL, or Off
(2.0 mmol/L to 15.0 mmol/L
in steps of 0.1 mmol/L, or Off)
When SpHb Precision is set to
1.0, the values are rounded
down.
When set to Off, alarm is
disabled.
Low Limit The Low Limit is
lower threshold
that triggers an
alarm.
7.0 g/dL
(4.0 mmol/L)
Off, or 1.0 g/dL to 23.5 g/dL in
steps of 0.1 g/dL
(Off, or 1.0 mmol/L to 14.5
mmol/L, in steps of 0.1
mmol/L)
When SpHb Precision is set to
1.0, values are rounded down.
When set to Off, alarm is
disabled.
Additional Settings for SpHb
From the Additional Settings screen, change any of the following options:
Options Description Factory Default
Settings
User Configurable
Settings
Averaging Time The length of time over
which the system
calculates the average of all
data points.
Medium Short, Medium, or Long
Calibration Provides an arterial or
venous value that displays
on the main screen.
Venous Arterial or Venous
Precision Allows the user to set the
decimal for SpHb.
0.1 0.1, 0.5, or 1.0 (whole
numbers)