User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 4: Operation
www.masimo.com 62 Masimo
SpO2 Settings
Access any of the following options:
Alarms for SpO2.
Additional Settings for SpO2 on page 63.
About Parameter Information on page 60.
Alarms for SpO2
From the Alarms screen, change any of the following options:
Options Description
Factory Default
Settings
Configurable Options
High Limit
High Limit is the upper
threshold that triggers an
alarm.
Off
2% to 99% in steps of
1%, or Off
When set to Off, alarm
is disabled
Low Limit
Low Limit is the lower
threshold that triggers an
alarm.
88%
1% to 98% in steps of
1%
Rapid Desat
Sets the Rapid Desat limit
threshold to the selected
amount below the Low
Alarm Limit. When SpO2
value falls below rapid desat
limit the audio and visual
alarm are immediately
triggered without respect to
the alarm delay.
-10% Off, -5%, or -10%
Alarm Delay
When an alarm condition is
met, this feature delays the
audible part of an alarm.
5 seconds 0, 5, 10, or 15 seconds
Adaptive
Threshold
Alarm (ATA)
ATA establishes
patient-specific limit
thresholds based upon the
baseline value of the
parameter.
See Adaptive Threshold
Alarm (ATA) Feature on
page 101.
Off Off or On