Radical-7 Chapter 2: Radical-7 Descriptions
www.masimo.com 40 Masimo
Item Connector Description
4 Power Entry module
The Power Entry module contains the input connector for
AC power and two fuses. The AC input provides power to
the system from the AC line. Always connect the
Radical-7 to the mains power for continuous operation
and/ or battery recharging. Note: Use the power cord as
the means to disconnect the instrument from the mains
power supply.
5 Equipotential
Ground connector
Use the Equipotential Ground connector for grounding.
Monitor Interface With SatShare
The Radical-7 has a unique SatShare interface that links to most existing multi-parameter
patient monitors through a SatShare cable.
Upgrades any approved and validated monitor to Masimo SET performance by
using the calculated SpO2 and pulse rate determined by Radical-7 to simulate an
ideal plethysmograph waveform, which is sent to the validated multi-parameter
patient monitor.
Connects into the SpO2 patient cable or SpO2 input connector of the
multi-parameter patient monitor.
See Setting Up and Using SatShare on page 44.