Radical-7 Chapter 2: Radical-7 Descriptions
www.masimo.com 39 Masimo
Standalone Back Panel
The following figure and corresponding text review the features of the Radical-7 Standalone.
Item Connector Description
1 Serial Output
connector
Use the Serial Output connector with a ferrite bead
installed to connect a serial instrument, including a
serial printer, a monitoring system or PC to the Radical-
7.
The data is provided in standard RS-232C format. All
external instrument connections to the Serial Output
connector must be IEC-60950 compliant.
2 Analog
Output/Nurse Call
connector
Use the Analog Output connector with a ferrite bead
installed to interface with an analog output instrument,
such as a chart recorder or Nurse Call system. All external
instrument connections to the Analog Output/Nurse Call
connector must be IEC-60950 compliant.
See Serial Interface Specifications on page 122.
3 SatShare Cable
connector
Use the SatShare Cable connector to connect a SatShare
cable to the SpO2 input connector of a multi-parameter
patient monitor. All external instrument connections to
the SatShare Cable Connector must be IEC-60601-1-1
compliant. SatShare cables are available to interface
with most major multi-parameter patient monitors.
Check the label on the SatShare cable and the SatShare
Directions for Use to ensure that the correct cable is used
for each type of patient monitor.
Visit www.masimo.com for the latest SatShare cables
and validated instruments.