Radical-7 Safety Information, Warnings, and Cautions
www.masimo.com 19 Masimo
Caution
Do not under any circumstances remove the grounding conductor from the
power plug.
Caution
Do not use extension cords or adapters of any type. The power cord and plug
must be intact and undamaged.
Caution
To ensure patient electrical isolation, connect only to other equipment with
electrically isolated circuits.
Caution
Do not connect to an electrical outlet controlled by a wall switch or dimmer.
All external instrument connections to the Analog Output/Nurse Call connector must be
IEC-60950 compliant.
It is recommended that the Radical-7 Handheld is docked to the Docking Station that is
attached to an AC power source when it is not in use to ensure that the battery remains fully
charged.
External instrument connections to the SatShare port must be IEC-60601-1 compliant.
Only use a SatShare cable that has a ferrite bead installed.
Use the power cord as the means to disconnect the instrument from the main power supply.
If the Radical-7 Handheld has not been used or charged within seven (7) days or more, then
recharge the battery prior to use.
The instrument must be configured to match your local power line frequency to allow for the
cancelation of noise introduced by fluorescent lights and other sources.
If there is any doubt about the integrity of the protective earth conductor arrangement,
operate the Radical-7 on internal battery power until the AC power supply protective
conductor is fully functional.
To prevent damage, do not soak or immerse the sensor in any liquid solution. Do not attempt
to sterilize by irradiation, steam, autoclave or any method other than ethylene oxide as
indicated.
Only SpO2 and pulse rate can be displayed on the multi-parameter monitor with SatShare.