Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements
www.masimo.com 146 Masimo
Adjust respiratory pause settings as necessary, default is 30 seconds (options are: 15,
20, 25, 30, 35, or 40 seconds).
Document patient events and time of event (based on computer clock). Include events
that may affect performance, these include:
Patient talking, picking at the sensor or nasal cannula, excessive movement,
ambient noise present, fans or air blowing at sensor
Inaccurate measurements may be caused by:
Excessive ambient or environmental noise (patient speaking, room noise)
Improper sensor placement
Cable disconnection
Movement, picking, or air blowing at sensor